User guide

GDM-360/397/398/461 User Manual
51
GDM-360
Range
Resolution
Accuracy
Input
Impedance
Fixed Value
Input
45~1kHz
>1kHz~3kHz
60mV
0.01mV
±(1.2%+5)
±(2.0%+5)
Around
>3000MΩ
1000V dc/
750Vrms ac
600mV
0.1mV
6V
0.001V
±(1.0%+3)
±(1.5%+5)
Around
10MΩ
60V
0.01V
600V
0.1V
750V
1V
±(1.2%+5)
±(3.0%+5)
GDM-360:
True RMS is applicable from 10% of range to 100% of range.
AC crest factor can be up to 3.0 except 750V where it is 1.5.
A residual reading of 10 digits with test leads shorted, will not affect
stated accuracy.