User guide

GDM-360/397/398/461 User Manual
57
GDM-360
Range
Resolution
Accuracy
Overload Protection
45~1kHz
1kHz~3kHz
600μA
0.1μA
±(1.2%+5)
±(1.5%+5)
Fuse 1: F600mA H 1000V,φ
6.35 x 31.8mm
6000μA
1μA
60mA
0.01mA
±(1.5%+5)
±(2.0%+5)
600mA
0.1mA
6A
0.001A
±(2.0%+5)
±(3.0%+5)
Fuse 2: F10A H 1000V,φ10.3 x
38.1mm
10A
0.01A
Remarks:
When ≤ 5A: Continuous measurement is allowed.
When > 5A: Continuous measurement for less than 10 seconds with intervals of
more than 15 minutes between measurements.
GDM-360:
True RMS is applicable from 10% of range to 100% for range.
A residual reading of 10 digits with test leads shorted, will not affect