For example, if [dBm] is pressed when measuring voltage in the max
mode, the maximum value is converted to dBm. To release the dBm
function, press [dBm] again. The dBm mode and AC+Hz mode are
not selected concurrently.
The standard reference impedance of the instrument is 600Ω.
5-6. AC+Hz measurements
The function can be selected only when ac range is selected. Press
[SHIFT], then press [AC+Hz], the secondary display shows the
frequency of the input signal that is higher than the sensitivity. The
frequency measurement does not depend on the max/min, rel, or hold
mode. In this mode, the reading rate of the DMM may be slower than
the normal state.
To release the AC+Hz function, press [AC+Hz] again. The dBm mode
and AC+Hz mode are not selected concurrently.
5-7 AC+DC measurements
The function can be selected only when voltage or current function is
selected. Press [AC+DC] button, the primary display shows the true
rms value of the input signal including the ac component and dc
component. In this mode, the reading rate of the DMM is slower than
the normal state.
To release the AC+DC function, press other function (voltage ac or
dc, current ac or dc, R,C, Continuity Beeper, Diode Test) key.
DUAL DISPLAY DIGITAL MULTIMETER
USER MANUAL
17
5-8 MAX/MIN measurements
The MAX/MIN mode causes the DMM to hold the lowest and highest
readings. Press [MAX/MIN] button to the MAX mode. The highest
will be displayed in continuous input. In the MAX mode, press
[MAX/MIN] button to the MIN mode. The lowest will be displayed in
continuous input. In the MIN mode, press [MAX/MIN] button to
release the MAX/MIN mode.
5-9. REL measurements
When the [REL] button is pressed, the meter stores the present reading
and displays subsequent measurements as the difference between the
measured value and the stored reading.
In the MAX/MIN mode, set [REL] button to the REL mode. The
maximum or minimum reading will become the relative base.
5-10. HOLD measurements
The HOLD mode that allows you to keep your eyes fixing on the
probes when taking measurements in difficult or hazardous
circumstances, then read the display when it is convenient and safe.
Press the [HOLD] button the last reading is held on the display in all
function. To release the HOLD function, press [HOLD] again.