User Manual

MAINTENANCE
135
7. The Channel1 calibration
will complete in less than 5
minutes.
Ch1 calibration 1/3
8. When finished, connect the
calibration signal to the
Channel 2 input and repeat
the procedure.
Done!!
9. When the calibration is complete the display
will go back to the previous state.
Probe Compensation
Procedure
1. Connect the probe between the Channel1 input
and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe voltage attenuation to x10.
2. Press the Utility key.
Utility
3. Press ProbeComp.