User Manual

GDS-1000A-U Series User Manual
22
6. Connect the probe between the Channel1 input
terminal and probe compensation signal output
(2Vp-p, 1kHz square wave).
7. Set the probe attenuation voltage to x10.
8. Press the Autoset key. A
square waveform will
appear in the center of the
display. For details on
Autoset, see page 48.
9. Press the Display key, then
Type and select the vector
waveform type.
10. Turn the adjustment point on the probe to
flatten the square waveform edge.