User Manual

MEASUREMENT
57
FRF
Time between:
Source 1 first rising edge and
Source 2 first falling edge
FFR
Time between:
Source 1 first falling edge and
Source 2 first rising edge
FFF
Time between:
Source 1 first falling edge and
Source 2 first falling edge
LRR
Time between:
Source 1 first rising edge and
Source 2 last rising edge
LRF
Time between:
Source 1 first rising edge and
Source 2 last falling edge
LFR
Time between:
Source 1 first falling edge and
Source 2 last rising edge
LFF
Time between:
Source 1 first falling edge and
Source 2 last falling edge
Automatic measurement gating
Background
Automatic measurements can be restricted to a
specific area (gating). When cursors are turned on,
the area between the cursors is used for automatic
measurements. When cursors are turned off,
measurements are derived from all the points that
are displayed on screen.
Turn gating on
1. Turn on cursors to enable
gated automatic
measurements.