Viewing waveforms in the X-Y mode
Background
The X-Y mode compares the voltage of Channel 1
and Channel 2 waveforms in a single display. This
mode is useful for observing the phase relationship
between the two waveforms.
Procedure
1. Connect the signals to
Channel 1 (X-axis) and
Channel 2 (Y-axis).
2. Make sure both Channel 1
and 2 are activated.
CH 1
CH 2
3. Press the Horizontal key.
MENU
4. Press XY. The display
shows two waveforms in X-
Y format; Channel 1 as X-
axis, Channel 2 as Y-axis.
Adjusting the X-Y
mode waveform
Horizontal position CH1 Position knob
Horizontal scale CH1 Volts/Div knob
Vertical position CH2 Position knob
Vertical scale CH2 Volts/Div knob
Example