User Manual

CONFIGURATION
99
Pressing Set to Minimum
sets the Holdoff time to the
minimum, 40ns.
Note: The holdoff function is automatically
disabled when the waveform update mode is
in Roll mode.
Configuring the edge trigger
Procedure
1. Press the Trigger menu key.
MENU
2. Press Type repeatedly to
select edge trigger.
3. Press Source repeatedly to
select the trigger source.
Range
Channel 1, 2, Line, Ext
4. Press Mode repeatedly to
select the Auto or Normal
trigger mode. To select the
single trigger mode, press
the Single key.
SINGLE
Range
Auto, Normal
5. Press Slope/coupling to enter
into the trigger slope and
coupling selection menu.
6. Press Slope repeatedly to
select the trigger slope,
rising or falling edge.
Range
Rising edge, falling edge