User manual

COMMAND DETAILS
117
LINe
AC Line
D0 to D15
Digital channels 1 to 15
Return parameter
Returns the trigger source.
Example
:TRIGger:SOURce CH1
Sets the trigger source to channel 1.
:TRIGger:COUPle
Set
Query
Description
Sets or queries the trigger coupling.
Note
Applicable for edge and delay triggers only.
Syntax
:TRIGger:COUPle {AC | DC | ?}
Parameter
AC
DC
Return parameter
Returns the trigger coupling.
Example
:TRIGger:COUPle AC
Sets the trigger coupling to AC.
:TRIGger:NREJ
Set
Query
Description
Sets or queries noise rejection status.
Syntax
:TRIGger:NREJ {OFF| ON| ?}
Parameter
OFF
Turns noise rejection off
ON
Turns noise rejection on
Return parameter
Returns the noise rejection status (ON, OFF).
Example
:TRIGger:NREJ ON
Turns noise rejection on.
:TRIGger:REJect
Set
Query
Description
Sets or queries the frequency rejection status.