165
:BUS1:PARallel:WIDth
Set
Query
Description
Sets or queries the number of bits for the parallel
bus.
Syntax
:BUS1:PARallel:WIDth {<NR1> | ? }
Parameter/Return
parameter
<NR1>
1~16
Example
:BUS1:PARallel:WIDth 16
Sets the width of the bus to 16 bits.
:BUS1:DISplay:FORMAt
Set
Query
Description
Sets or queries the display format for the bus,
either binary or hexadecimal.
Syntax
:BUS1:DISplay:FORMAt { BINary | HEXadecimal | ? }
Parameter/Return
parameter
BINary
Binary format
HEXadecimal
Hexadecimal format
Example
: BUS1:DISplay:FORMAt BINary
Sets the display format to binary.