GDS-2000A Programming Manual
64
Syntax
:CURSor:HDELta {?}
Return Parameter
<NR3>
Returns the distance between two
horizontal cursors.
Example
:CURSor:HDELta?
5.0E-9
Returns the horizontal delta as 5ns.
:CURSor:V1Position
Set
Query
Description
Sets the first vertical cursor (V1) position.
Syntax
:CURSor:V1Position {<NRf>| ?}
Parameter
<NRf>
Vertical position. Depends on the
vertical scale.
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:V1Position 1.6E -1
Sets the V1 cursor position to 160mA.
:CURSor:V2Position
Set
Query
Description
Sets the first vertical cursor (V2) position.
Syntax
:CURSor:V2Position {<NRf> | ?}
Parameter
<NRf>
Vertical position. Depends on the
vertical scale.
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:V2Position 1.1E-1
Sets the V2 cursor position to 110mA.
:CURSor:VDELta
Query
Description
Returns the delta of V1 and V2.
Syntax
:CURSor:VDELta {?}