Datasheet
MEASUREMENT
67
Change modes
1. Choose between Fit Screen Mode
and AC Priority Mode from the
bottom menu.
2. Press the Autoset key again to use
Autoset in the new mode.
Autoset
Fit Screen Mode
AC Priority
Limitation
Autoset does not work in the following situations.
Input signal frequency is less than 20Hz
Input signal amplitude is less than 30mV
Note
The Autoset key (page 66) does NOT automatically
activate the channels to which input signals are
connected.