GDS-2000A Programming Manual
42
Example
:ACQuire1:STATe?
0
Returns 0. The channel 1’s raw data is not ready.
Note: If the oscilloscope changes the acquisition status
from STOP to RUN, the status will be reset as zero.
:ACQuire:INTERpolation
Set
Query
Description
Selects or returns the interpolation mode.
Syntax
:ACQuire:INTERpolation {ET | SINC | ?}
Parameter/Return
parameter
ET
Set the Equivalent Time interpolation.
SINC
Sets to SIN(X)/X interpolation
Example
:ACQuire:INTERpolation ET
Sets the scope to ET interpolation.
:ACQuire:FILTer
Set
Query
Description
Sets the normalized cut-off frequency to the
nearest set. [0.02 : 0.02~0.98]
Syntax
:ACQuire:FILTer {OFF | <NR3> | ?}
Parameter/Return
parameter
OFF
Turns the digital filter off.
<NR3>
0.02,0.04~0.98.
Example
:ACQuire:FILTer OFF
Turns the digital filter off.
:ACQuire:RECOrdlength
Set
Query
Description
Sets or queries the record length. Please see the
user manual for full details.
Syntax
:ACQuire:RECOrdlength {AUTo | SHORT | ?}
Parameter/Return
AUTo
Auto record length.