:CHANnel<X>:PROBe:RATio
Set
Query
Description
Sets or returns the probe attenuation factor.
Syntax
:CHANnel<X>:PROBe:RATio { <NRf> | ?}
Related
Commands
:CHANnel<X>:PROBe:TYPe
Parameter
<X>
Channel 1, 2, 3, 4
<NRf>
Probe attenuation factor.
Return parameter
<NR3>
Returns the probe factor.
Example
:CHANnel1:PROBe:RATio 1.00E+0
Sets the Channel 1 probe attenuation factor to 1x
:CHANnel<X>:PROBe:TYPe
Set
Query
Description
Sets or returns the probe type (voltage/current).
Syntax
:CHANnel<X>:PROBe:TYPe { VOLTage | CURRent | ?}
Related
Commands
:CHANnel<X>:PROBe:RATio
Parameter
<X>
Channel 1, 2, 3, 4
VOLTage
Voltage
CURRent
Current
Return parameter
Returns the probe type.
Example
:CHANnel1:PROBe:TYPe VOLTage
Sets the Channel 1 probe type to voltage.
:CHANnel<X>:SCALe
Set
Query
Description
Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Note the probe attenuation factor should be set
before the scale.