User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Ethernet Commands
- Time Commands
- Bus Decode Commands
- Mark Commands
- Search Commands
- Digital Commands
- Label Commands
- Utility Commands
- APPENDX
- INDEX
GDS-2000A Programming Manual
88
Syntax
:MEASure:CMEan{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the cyclic mean.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:CMEan?
9.480E-01
Selects Channel 1, and then measures the mean
value of the first period.
:MEASure:HIGH
Query
Description
Returns the high voltage/current.
Syntax
:MEASure:HIGH{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the high value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:HIGH?
3.68E-3
Selects Channel 1, and then measures the high
voltage/current.