Datasheet

GDS-2000A Series User Manual
246
FAQ
I connected the signal but it does not appear on the display.
I want to remove the (Measurement result / FFT result / Help
contents) from the display.
The waveform does not update (frozen).
The probe waveform is distorted.
Autoset does not catch the signal well.
I can’t save files to the internal memory.
The display image printout is too dark on the background.
The date and time settings are not correct.
The accuracy does not match the specification.
I connected the signal but it does not appear on the display.
Make sure you have activated the channel by pressing the Channel
key (the channel key lights up).
I want to remove the (Measurement result / FFT result /
Help contents) from the display.
To clear automatic measurement results, press the Measure key,
select Remove Measurement and choose Remove All. See page 78.
To clear individual measurements from the screen, press the
Measure key, select Display All and choose Off. See page 80.
To clear the FFT result, press the Math key twice. See page 93 for
details.
To clear Help result, press the Help key again. See page 63 for