PEL-3000 Programming Manual
110
Return parameter
<string>
Returns the normal
sequence settings in the
following order:
Start:<NR1>, Seq No:
<NR1>, Memo:<string>,
Mode:CC|CR|CV|CP,
Range:
ILVL|IMVL|IHVL|ILVH|IMV
H|IHVH, Loop:<NR1>,
Last Load: OFF|ON,
Last:<NR2>, Chain:<NR1>
Example
:NSEQuence: 1,1,”hi”,CC,ILVL,1,ON,0.000,2
Sets the normal sequence configuration to that shown
below.
Query example
:NSEQuence:?
> Start:1, Seq No:1, Memo: HI, Mode:CC,
Range:IHVL, Loop:1, Last Load:ON, Last:0.000,
Chain:2
Returns the normal sequence settings.
:NSEQuence:STARt
Set
Query
Description
Sets or queries the starting sequence number.
Syntax
:NSEQuence:STARt {<NR1>}
Query Syntax
:NSEQuence:STARt?
Parameter
<NR1>
(1~10) sequence number
Return Parameter
<string>
Start:<NR1>
Example
:NSEQuence:STARt 1
Sets the starting sequence number to 1.
Query example
:NSEQuence:STARt?
>Start:1
Returns the starting sequence number (1).