User manual

PEL-3000 Programming Manual
52
Query example
:GNG?
>GO
Returns the Go-NoGo test result.
[:CONFigure]:GNG:H
Set
Query
Description
Sets or queries the high voltage/current limit
value. If the entry mode is set to value, the high
voltage/current limit value units are in
volts/amps. If the entry mode is set to percent, the
high voltage/current limit value units are in
percent.
Syntax
[:CONFigure]:GNG:H <NRf>
Query Syntax
[:CONFigure]:GNG:H?
Parameter
<NRf>
Sets the high
voltage/current limit value
in volts/amps or in
percent.
Return parameter
<NR2>
Returns the high
voltage/current limit value
in volts/amps or as %.
Example
:GNG:H 100.0
Sets the high voltage limit value to 100%.
Query example
:GNG:H?
>100.0
Returns the high voltage limit value as 100.0%.
[:CONFigure]:GNG:L
Set
Query
Description
Sets or queries the low voltage/current limit value.
If the entry mode is set to value, the low
voltage/current limit value units are in
volts/amps. If the entry mode is set to percent, the
low voltage/current limit value units are in
percent.
Syntax
[:CONFigure]:GNG:L <NRf>