COMMNAND OVERVIEW
53
Query Syntax
[:CONFigure]:GNG:L?
Parameter
<NRf>
Sets the low
voltage/current limit value
in volts/amps or in
percent.
Return parameter
<NR2>
Returns the low
voltage/current limit value
in volts/amps or as %.
Example
:GNG:L 10.0
Sets the low voltage limit value to 10%.
Query example
:GNG:L?
>10.0
Returns the low voltage limit value as 10.0%.
[:CONFigure]:GNG:C
Set
Query
Description
Sets or queries the center voltage/current limit
value. The center voltage limit value is used as the
center reference value when the entry mode is set
to percent ([:CONFigure]:GNG:Mode=PERCent).
Syntax
[:CONFigure]:GNG:C <NRf>
Query Syntax
[:CONFigure]:GNG:C?
Parameter
<NRf>
Sets the center
voltage/current limit value
in volts/amps.
Return parameter
<NR2>
Returns the center
voltage/current limit value
in volts/amps.
Example
:GNG:C 10.0
Sets the center voltage/current limit value to 10V or A.