PEL-3000 Programming Manual
58
Parameter
<NRf>[ A ]
Step resolution.
MINIMUM or MIN
Minimum step resolution
MAXIMUM or MAX
Maximum step resolution
Return parameter
<ASCII string>
Returns the range and the
step resolution.
Example
:STEP:CCL 0.02mA
Sets the step resolution to 0.02mA.
Query example
:STEP:CCL?
> CCL:0.00002
Returns the step resolution (0.00002A).
[:CONFigure]:STEP:CC
Set
Query
Description
Returns the step resolution for each CC Mode
range as a string.
Query Syntax
[:CONFigure]:STEP:CC?
Return parameter
<ASCII string>
Returns the CCH, CCM
and CCL step resolution
settings.
Query example
:STEP:CC?
>CCH:0.002, CCM:0.0002, CCL:0.00002
Returns the CC mode step resolution for each range.
[:CONFigure]:STEP:CRH
Set
Query
Description
Configures the step resolution for CR High Range.
Note: The step resolution setting will be
automatically rounded to the closest multiple of
the base resolution.
Syntax
[:CONFigure]:STEP:CRH {<NRf>[ MHO ] | MINimum
| MAXimum }
Query Syntax
[:CONFigure]:STEP:CRH?
Parameter
<NRf>[ MHO ]
Step resolution. [MHO]
MINIMUM or MIN
Minimum step resolution
MAXIMUM or MAX
Maximum step resolution