User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
COMMNAND OVERVIEW
73
Syntax
:CURRent:L2 {<NRf>[ A ] | MINimum | MAXimum }
Query Syntax
:CURRent:L2?
Parameter
<NRf>[A]
“Level2” current value
MINIMUM or MIN
Minimum current level
MAXIMUM or MAX
Maximum current level
Return parameter
<NR2>
Returns the “Level2”
current value.
Example
:CURRent:L2 MIN
Sets the current value to the minimum.
Query example
:CURRent:L2?
>1.0A
Returns the “Level2” current setting.
:CURRent:SET
Set
Query
Description
Sets or queries the CC dynamic mode Set value
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.
Syntax
:CURRent:SET{<NRf>[ A ] | MINimum | MAXimum }
Query Syntax
:CURRent:SET?
Related
Commands
:CURRent:LEVel
Parameter
<NRf>[A]
“Set” current value
MINIMUM or MIN
Minimum current level
MAXIMUM or MAX
Maximum current level
Return parameter
<NR2>
Returns the “Set” current
value.
Example
:CURRent:SET MIN
Sets the current value to the minimum.
Query example
:CURRent:SET?
>1.0A
Returns the “Set” current setting.