User manual

PEL-3000 Programming Manual
74
:CURRent:LEVel
Set
Query
Description
Sets or queries the CC mode % level (percentage of
the Set current value) when the dynamic mode
units are set to Percent.
Syntax
:CURRent:LEVel {<NRf>| MINimum | MAXimum }
Query Syntax
:CURRent:LEVel?
Related
Commands
:CURRent:SET
Parameter
<NRf>
% level” current value
MINIMUM or MIN
Minimum % current level
MAXIMUM or MAX
Maximum % current level
Return parameter
<NR2>
Returns the “% level
current value.
Example
:CURRent:LEVel MIN
Sets the % level current value to the minimum.
Query example
:CURRent:LEVel?
>50
Returns the “% Levelas 50 percent.
:CURRent:RISE
Set
Query
Description
Sets the rising current slew rate for CC dynamic
mode.
Syntax
:CURRent:RISE {<NRf> | MINimum | MAXimum }
Query Syntax
:CURRent:RISE?
Related
Commands
:CURRent:FALL
Parameter
<NRf>
Rising current slew rate
MINIMUM or MIN
Minimum slew rate
MAXIMUM or MAX
Maximum slew rate
Return parameter
<NR2>
Returns the rising current
slew rate in mA/us.