126
Parameter
Parameters in order:
Description
<NR2>
Start_Value
<NR2>
End_Value
<NR1>
Start_Step
<NR1>
End_Step
Return parameter
<string>
Returns the settings for
the Fill Edit for Fast
Sequence menu:
Start Value:<NR2>, End
Value:<NR2>, Start Step:
<NR1>, End Step:<NR1>
Example
:FSEQuence:FILL 0.1,0.5,1,159
Fills the values of step 1 to step 159 gradually from
0.1A to 0.5A.
Query example
:FSEQuence:FILL?
>Start Value:0.10000, End Value:0.50000, Start Step:1,
End Step:159
:FSEQuence:SAVE
Set
Description
Saves the data points for the fast sequence.
Syntax
:FSEQuence:SAVE
Example
:FSEQuence:SAVE