COMMNAND OVERVIEW
161
-111 Header separator error
A character which is not a legal
header separator was
encountered while parsing the
header.
-112 Program mnemonic too long
The header contains more than
twelve characters.
-113 Undefined header
The header is syntactically
correct, but it is undefined for this
specific device.
-114 Header suffix out of range
The value of a numeric suffix
attached to a program mnemonic.
-115 Unexpected number of
parameters
The number of parameters
received does not correspond to
the number of parameters
expected.
-120 Numeric data error
This error is generated when
parsing a data element which
appears to be numeric, including
the nondecimal numeric types.
-121 Invalid character in number
An invalid character for the data
type being parsed was
encountered.
-123 Exponent too large
The magnitude of the exponent
was larger than 32000.
-124 Too many digits
The mantissa of a decimal
numeric data element contained
more than 255 digits excluding
leading zeros.
-128 Numeric data not allowed
A legal numeric data element was
received, but the device does not
accept one in this position for the
header
-130 Suffix error
This error, as well as errors -131
through -139, are generated when
parsing a suffix.