PEL-3000 Programming Manual
46
Example
:VDELay 1.5 ms
Sets the delay time to 1.5ms.
:VDELay 0.0015 s
Sets the delay time to 1.5ms.
Query example
:VDELay?
>0.0015
The delay time is 1.5ms.
:CONFigure:RESPonse
Set
Query
Description
Sets or queries the response speed. This is the
equivalent to the CC, CR and CP Mode.
Syntax
:CONFigure:RESPonse {<NR2> | MINimum |
MAXimum }
Query Syntax
:CONFigure:RESPonse?
Parameter
<NR2>
0.1, 0.2, 0.5, 1.0 (for
CC ,CR or CP mode only)
MINimum
Minimum response
speed
MAXimum
Maximum response
speed
Return Parameter
<NR2>
0.1, 0.2, 0.5, 1.0
Example
:CONFigure:RESPonse MAX
Sets the response to the maximum.
Query example
:CONFigure:RESPonse?
>1.0
Returns the response.
[:CONFigure]:CNTime
Set
Query
Description
Turns the Count Time timer function on or off.
Syntax
[:CONFigure]:CNTime {<Boolean> |OFF | ON }
Query Syntax
[:CONFigure]:CNTime?