User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
PEL-3000 Programming Manual
50
TOGGLE or TOGG
Sets the button
configuration to toggle
Return parameter
Toggle
Toggle
Hold
Hold
Example
:CONFigure:SHORt TOGGle
Sets the Short key configuration to toggle.
Query example
:CONFigure:SHORt?
>Toggle
The Short key is configured to toggle.
[:CONFigure]:GNG:SPECtest
Set
Query
Description
Enables/Disables Go-NoGo testing (SPEC test =
ON/SPEC test = OFF).
Syntax
[:CONFigure]:GNG:SPECtest {<Boolean>|OFF | ON }
Query Syntax
[:CONFigure]:GNG:SPECtest?
Parameter/
Return parameter
OFF or 0
SPEC test = OFF
ON or 1
SPEC test = ON
Example
:GNG:SPECtest ON
Turns Go-NoGo testing on.
Query example
:GNG:SPECtest?
>OFF
Indicates that Go-NoGo testing is off.
[:CONFigure]:GNG:DTIMe
Set
Query
Description
Sets or queries the Go-NoGo delay time.
Syntax
[:CONFigure]:GNG:DTIMe {<NRf>[s] | MINimum
|MAXimum }
Query Syntax
[:CONFigure]:GNG:DTIMe?
Parameter
<NRf>[s]
Sets the Go-NoGo delay
time in seconds (0.0~1.0)
with 0.1 second resolution.
MINIMUM or MIN
Minimum delay time
MAXimum or MAX
Maximum delay time