User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
COMMNAND OVERVIEW
69
Parameter
FAST
Fast response
SLOW
Slow response
Return parameter
<ASCII string>
Returns CV response
speed as a string.
Example
:RESPonse FAST
Sets the CV response to fast.
Query example
:RESPonse?
> CV mode response:FAST.
[:MODE]:DYNamic
Set
Query
Description
Sets or queries the switching mode. Dynamic is the
default mode.
Syntax
:MODE:DYNamic{ DYNamic | STATic }
Query Syntax
:MODE:DYNamic?
Parameter
DYNAMIC or DYN
Set to Dynamic mode
STATIC or STAT
Set to Static mode
Return parameter
<ASCII string>
Returns the switching
mode.
Example
:MODE:DYNamic DYN?
Set the switching mode to dynamic
Query example
:MODE:DYNamic?
> Dynamic, ;Dynamic Level:Value, Dynamic
Time:T1/T2
The switching mode is set to dynamic mode.