PEL-3000 Programming Manual
72
Return parameter
<NR2>
Returns the slew rate in
mA/us.
Example
:CURRent:SRATe MIN
Sets the slew rate to the minimum.
Query example
:CURRent:SRATe?
>5.0
Returns the slew rate as 5.0mA/us.
:CURRent:L1
Set
Query
Description
Sets or queries the CC mode “Level 1” current.
This command is only applicable to dynamic
modes.
Note: A different current value can be set for each
range (High/Mid/Low).
Syntax
:CURRent:L1 {<NRf>[ A ] | MINimum | MAXimum }
Query Syntax
:CURRent:L1?
Parameter
<NRf>[A]
“Level1” current value
MINIMUM or MIN
Minimum current level
MAXIMUM or MAX
Maximum current level
Return parameter
<NR2>
Returns the “Level1”
current value.
Example
:CURRent:L1 MIN
Sets the current value to the minimum.
Query example
:CURRent:L1?
>1.0A
Returns the “Level1” current setting.
:CURRent:L2
Set
Query
Description
Sets or queries the CC mode “Level2” current. This
command is only applicable to dynamic modes.
Note: A different current value can be set for each
range (High/Mid/Low).