User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
COMMNAND OVERVIEW
75
Example
:CURRent:RISE MIN
Sets the rising slew rate to the minimum.
Query example
:CURRent:RISE?
>5000
Returns the rising slew rate as 5000mA/us.
:CURRent:FALL
Set
Query
Description
Sets the falling current slew rate for CC dynamic
mode.
Syntax
:CURRent:FALL {<NRf> | MINimum | MAXimum }
Query Syntax
:CURRent:FALL?
Related
Commands
:CURRent:FALL
Parameter
<NRf>
Falling current slew rate
MINIMUM or MIN
Minimum slew rate
MAXIMUM or MAX
Maximum slew rate
Return parameter
<NR2>
Returns the falling current
slew rate in mA/us.
Example
:CURRent:FALL MIN
Sets the falling slew rate to the minimum.
Query example
:CURRent:FALL?
>5000
Returns the falling slew rate as 5000mA/us.
:CURRent:T1
Set
Query
Description
Sets the T1 timer settings for the CC dynamic
mode.
Syntax
:CURRent:T1 {<NRf>[s] | MINimum | MAXimum}
Query Syntax
:CURRent:T1?
Related
Commands
:CURRent:T2