User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
PEL-3000 Programming Manual
76
Parameter
<NRf>[s]
T1 timer setting in
seconds
MINIMUM or MIN
Minimum time
MAXIMUM or MAX
Maximum time
Return parameter
<NR2>
Returns the T1 timer time
in seconds.
Example
:CURRent:T1 0.200
Sets the T1 timer time to 200ms.
Query example
:CURRent:T1?
>0.200
Returns the T1 timer time as 200ms.
:CURRent:T2
Set
Query
Description
Sets the T2 timer settings for the CC dynamic
mode.
Syntax
:CURRent:T2 {<NRf>[s] | MINimum | MAXimum}
Query Syntax
:CURRent:T2?
Related
Commands
:CURRent:T1
Parameter
<NRf>[s]
T2 timer setting in
seconds
MINIMUM or MIN
Minimum time
MAXIMUM or MAX
Maximum time
Return parameter
<NR2>
Returns the T2 timer time
in secondes.
Example
:CURRent:T2 0.200
Sets the T2 timer time to 200ms.
Query example
:CURRent:T2?
>0.200
Returns the T2 timer time as 200ms.