COMMNAND OVERVIEW
77
:CURRent:FREQuency
Set
Query
Description
Sets the CC dynamic mode switching frequency.
This command along with the “:CURRent:DUTY”
command is used when the dynamic time setting
is set to “Freq./Duty”.
This command is only applicable for dynamic
mode.
Syntax
:CURRent:FREQuency {<NRf> | MINimum |
MAXimum}
Query Syntax
:CURRent:FREQuency?
Related
Commands
:CURRent:DUTY
Parameter
<NRf>
Sets the switching
frequency in hertz.
MINIMUM or MIN
Minimum frequency
MAXIMUM or MAX
Maximum frequency
Return parameter
<NR2>
Returns the frequency in
hertz.
Example
:CURRent:FREQuency 60
Sets frequency to 60Hz.
Query example
:CURRent:FREQuency?
>60
Returns the switching frequency (60Hz).
:CURRent:DUTY
Set
Query
Description
Sets the positive duty cycle for the CC dynamic
mode switching frequency. This command along
with the “:CURRent:FREQuency” command is
used when the dynamic time setting is set to
“Freq./Duty”.
This command is only applicable for dynamic
mode.