COMMNAND OVERVIEW
83
Syntax
:RESistance:SET {<NRf> | MINimum | MAXimum }
Query Syntax
:RESistance:SET?
Related
Commands
:RESistance:LEVel
Parameter
<NRf>
“Set” conductance value
MINIMUM or MIN
Minimum conductance
level
MAXIMUM or MAX
Maximum conductance
level
Return parameter
<NR2>
Returns the “Set”
conductance value.
Example
:RESistance:SET MIN
Sets the conductance to the minimum.
Query example
:RESistance:SET?
>.200
Returns the “Set” conductance setting in MHO.
:RESistance:LEVel
Set
Query
Description
Sets or queries the CR mode % level (percentage of
the Set conductance value) when the dynamic
mode units are set to Percent.
Syntax
:RESistance:LEVel {<NRf> | MINimum | MAXimum }
Query Syntax
:RESistance:LEVel?
Related
Commands
:RESistance:SET
Parameter
<NRf>
“% level” conductance
value
MINIMUM or MIN
Minimum % conductance
level
MAXIMUM or MAX
Maximum % conductance
level
Return parameter
<NR2>
Returns the “% level”
conductance value.