COMMNAND OVERVIEW
91
Power Subsystem Commands
:POWer[:VA] ......................................................................... 91
:POWer:VB............................................................................ 92
:POWer:L1 ............................................................................. 92
:POWer:L2 ............................................................................. 93
:POWer:SET ......................................................................... 93
:POWer:LEVel ...................................................................... 94
:POWer:T1 ............................................................................ 95
:POWer:T2 ............................................................................ 95
:POWer:FREQuency ........................................................... 96
:POWer:DUTY ..................................................................... 97
:POWer[:VA]
Set
Query
Description
Sets or queries the CP mode “A Value” power.
This command is applicable to static mode only.
Note: A different “A Value” can be set for different
current ranges.
Note: The [:VA] node can only be omitted when in
static mode.
Syntax
:POWer[:VA] {<NRf>[ W ] | MINimum | MAXimum }
Query Syntax
:POWer[:VA]?
Parameter
<NRf>[ W ]
“A Value” power value
MINIMUM or MIN
Minimum power level
MAXIMUM or MAX
Maximum power level
Return parameter
<NR2>
Returns the “A Value”
power value.
Example
:POWer:VA MIN
Sets the power value to the minimum.
Query example
:POWer:VA?
>10W