95
:POWer:T1
Set
Query
Description
Sets the T1 timer settings for the CP dynamic
mode.
Syntax
:POWer:T1 {<NRf>[s] | MINimum | MAXimum}
Query Syntax
:POWer:T1?
Related
Commands
:POWer:T2
Parameter
<NRf>[s]
T1 timer setting in
seconds
MINIMUM or MIN
Minimum time
MAXIMUM or MAX
Maximum time
Return parameter
<NR2>
Returns the T1 timer time
in s.
Example
:POWer:T1 0.200
Sets the T1 timer time to 200ms.
Query example
:POWer:T1?
>0.200
Returns the T1 timer time as 200ms.
:POWer:T2
Set
Query
Description
Sets the T2 timer settings for the CP dynamic
mode.
Syntax
:POWer:T2 {<NRf>[s]| MINimum | MAXimum}
Query Syntax
:POWer:T2?
Related
Commands
:POWer:T1
Parameter
<NRf>[s]
T2 timer setting in
seconds
MINIMUM or MIN
Minimum time
MAXIMUM or MAX
Maximum time
Return parameter
<NR2>
Returns the T2 timer time
in ms.