User manual

REMOTE CONTROL
109
CV (Constant voltage mode)
Indicates if the PSU is in CV
mode.
8
256
CC (Constant current mode)
Indicates if the PSU is in CC
mode.
10
1024
OND (Output ON Delay)
Indicates if Output ON delay time
is active
11
2048
OFD (Output OFF Delay)
Indicates if Output OFF delay
time is active
12
4096
PR (Program Running)
Indicates if a Test is running
14
16384
Condition
Register
The Operation Status Condition Register
indicates the operating status of the power
supply. If a bit is set in the Condition register, it
indicates that the event is true. Reading the
condition register does not change the state of
the condition register.
PTR/NTR Filters
The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will set the corresponding bit in
the Event Registers. Use the Positive transition
filter to view events that change from false to
positive, and use the negative transition filter to
view events that change from positive to
negative.
Positive Transition
01
Negative Transition
10