User manual
PSU Programming Manual
86
[:SOURce]:VOLTage[0]:SLEW:RISing
Set
Query
Description
Sets or queries the rising voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
Query Syntax
[:SOURce]:VOLTage[0]:SLEW:RISing
{<NR2>(V)|MINimum|MAXimum}
[:SOURce]:VOLTage[0]:SLEW:RISing?
Parameter
<NR2>
0 ~ Vrated/100msec
MIN
Minimum rising voltage slew rate
MAX
Maximum rising voltage slew rate
Return parameter
<NR2>
Returns the slew rate in A/ms.
Example
SOUR:VOLT:SLEW:RIS MAX
Sets the rising voltage slew rate to its maximum.
[:SOURce]:VOLTage[0]:SLEW:FALLing
Set
Query
Description
Sets or queries the falling voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
[:SOURce]:VOLTage[0]:SLEW:FALLing
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage[0]:SLEW:FALLing?
Parameter
<NR2>
0 ~ Vrated/100msec
MIN
Minimum voltage falling slew rate
MAX
Maximum voltage falling slew rate
Return parameter
<NR2>
Returns the voltage slew rate in A/ms
Example
SOUR:VOLT:SLEW:FALL MIN
Sets the falling voltage slew rate to its minimum.