User manual

REMOTE CONTROL
89
3
External resistance control; 10kΩ = Io min,
0kΩ = Io max.
:SYSTem:CONFigure:VOLTage:CONTrol
Set
Query
Description
Sets or queries the CV control mode (local control
(panel), external voltage control, external
resistance control). This setting is applied only
after the unit is reset.
Syntax
Query Syntax
:SYSTem:CONFigure:VOLTage:CONTrol { <NR1> }
:SYSTem:CONFigure:VOLTage:CONTrol?
Parameter/Return
<NR1>
Description
0
Local (Panel) control
1
External voltage control
2
External resistance control; 10kΩ = Io max,
0kΩ = Io min.
3
External resistance control; 10kΩ = Io min,
0kΩ = Io max.
:SYSTem:CONFigure:OUTPut:PON[:STATe]
Set
Query
Description
Sets the output state at power-on. This is the
equivalent to the F-92 (Output Status when Power
ON) power on configuration settings.
Syntax
Return Syntax
:SYSTem:CONFigure:OUTPut:PON[:STATe]
{<NR1>|SAFE|FORCe |AUTO}
:SYSTem:CONFigure:OUTPut:PON[:STATe]?
Parameter
SAFE | 0
The PSU turns on in the same state the unit
was in prior to the previous shut down. The
output is set to off (default).
FORCe | 1
The PSU turns on in the same state the unit
was in prior to the previous shut down. The
output is set to on.
AUTO | 2
The PSU turns on in the same state the unit
was in prior to the previous shut down, but
with the same output on/off setting.
Return parameter
0
The power on output setting is “SAFE”.
1
The power on output setting is “FORCe”.
2
The power on output setting is “AUTO”.