8544Maintenance.fm Maintenance
Section 5 Maintenance
DANGER
Only qualified personnel should conduct the installation tasks described in this
section of the manual
5.1 Cleaning the Sampler
5.1.1 Cleaning the Sampler Housing
The interior and exterior sampler housing may be cleaned with a damp
sponge and mild detergent. Do not use abrasive cleaners.
5.1.2 Cleaning the Sample Bottles
Clean the bottles using a brush and water with mild detergent, followed by a
fresh water rinse and a diluted water rinse. Glass bottles may also be
autoclaved.
5.1.3 Cleaning the Intake Tubing and Pump Tubing
Clean the tubing by pumping water with mild detergent or other cleaning
solution through the tubing, using the peristaltic pump in Manual Pump Mode.
Press the
MANUAL MODE soft key.
5.1.4 No Lubrication Required
The 900 Composite Sampler does not require routine lubrication. Do not
lubricate the pump tube with petroleum jelly, silicone grease, oil, or any other
lubricant, as this will substantially reduce the life of the pump tube and roller
assembly.
5.2 Pump Tubing Maintenance
5.3 Upgrades, Repairs, General Maintenance
5.4 Electrostatic Discharge (ESD) Considerations
5.5 Internal Maintenance Items
The following item requires access to the inside of the case for repairs:
Internal desiccant module
RAM memory battery
5.6 Removing and Opening the Controller
CAUTION
Always power the unit OFF and
then disconnect all cables from
the 900 Composite sampler
before removing the controller.
To remove the controller from the controller center section:
1. Turn the unit off by pressing the OFF key.
2. Disconnect and remove all cables to the controller housing, including the
distributor/full bottle shutoff cable.
3. Disconnect all tubing.