Product Manual

C1000 - User Guide Rev: 9SEP2011 Page: 21
The lift unit also has an Emergency Shut-off feature that allows the operator
to shut the power to the lift unit completely off. By pulling down once on the
RED emergency lowering cord, located on the underside of the lift unit, the
lift will immediately stop and all its functions will be disabled. The unit will
beep once and all power to the lift will be turned off. The ON Indicator light
and display will turn off, and the Emergency Shut-off button located inside
the lift case will pop out. Should this feature be used, contact your local
authorized dealer immediately. The lift unit must be inspected prior to
restoring to use. In order to restore power back to the lift unit, the white
plastic tab that popped out when the cord was pulled, can be easily pressed
back into the lift case by use of your finger.
Emergency Stopping
In the event that the DOWN button on the hand control does not function,
or in power failure situations, the person may be lowered by pulling down
and holding the RED emergency lowering cord located on the underside of
the lift unit. Continue to pull down until the person is safely lowered to the
desired position. The unit will beep as you continue to pull down on the
cord and will continue beeping until the cord is released after the desired
lowering has been achieved. Please refer to figure 13A.
IMPORTANT: The Emergency Lowering button does not provide a rais-
ing function. The failure of any of the lowering devices should be reported
to Prism Medical or your authorized dealer immediately.
Figure 12A - Pull down on the
RED CORD once to stop the lift.
The unit will beep once and all
power to the lift will be turned off.
Emergency Lowering
Figure 13A - Pull down and hold
until the person is safely lowered to
the desired position.
Figure 12B – Return power to the lift
unit by pressing in the plastic tab
Once the RED Emergency Stop/
Lowering Cord is released the lift
unit will need to be reset in order to
operate again. This can be achieved
buy pressing the plastic tab located
at the end of the RED CORD, back
into the lift unit. Then, simply press
any button on the hand control to
resume power. Please refer to Fig-
ure 12B.
Pull
Down
and
Hold
Pull
Down
Once