Product Manual

C1000 - User Guide Rev: 9SEP2011 Page: 22
Emergency Manual Raising or Lowering
Caution: A proper safety ladder or stool may be required in order to remove the cover of the device.
Use extreme caution if this is required. Should you have any concerns or questions contact your local
authorized Prism Medical dealer.
Caution: The manual emergency raising and lowering system should be used only if the lowering
procedures described in the previous section of the manual do not work, or, if the emergency raising
function is required. Should you have any concerns or questions contact your local authorized Prism
Medical dealer.
Figure 14A– Locate Allen Key supplied
with the lift.
Step 1) Locate the
Allen key in the plas-
tic bag containing the
owners manual, clev-
is pin & ring, and
hand control.
Step 2) Locate a small
circular clip on one
side of the lift. Refer
to Figure 14B. Re-
move this clip and pro-
ceed to Step 3.
Figure 14B - Remove clip from side of the
lift.
Step 3) Insert the Allen key into the hole dis-
covered after removing the clip and turn the key
with your fingers. Refer to Figure 14C. Turn
the clockwise to lower the lift strap and counter
clockwise to raise.
Figure 14C - Turn the Allen Key to lift/ low-
er the lift strap.
Caution: DO NOT use the lift after the manual lowering mechanism has been used. The lift must be reset
by a qualified lift technician after use. contact your local authorized Prism Medical dealer.