General Inspection and Maintenance
A) Each Use - To be completed by User
Prior to each use the
C1000 lift and associated track, accessories and sling (s), must be visually inspected. Refer to the
accessory and sling user guides for specific details regarding their inspection.
Should any of the these items fail the inspection do not use the lift
Contact your local authorized dealer for service.
Visually check for the following:
The lift lifting tape shows NO signs of fraying or breaking along its entire length.
The stitching on the lift lifting tape where it connects to the carry bar shows NO signs of fraying, or
breaking.
The sling (s) that will be used shows NO signs of unusual wear and tear. The straps of the sling that connect
to the carry bar of the lift show NO signs of fraying or breaking. Refer to specific sling instructions.
The carabiner hook shall be visually inspected before and after each lift for damages.
The airline tube that connects the hand control to the lift is not kinked, twisted, knotted, cut or damaged.
All the functions on the hand control work correctly (e.g. UP/DOWN/.LEFT/RIGHT, etc..).
The brackets that hold the track in place on the ceiling are secure and do not move or appear loose.
There are not cuts, dents or sharp edges on the carry bar that may damage the straps of the sling.
The lift has no unusual sounds when the carry bar is moved UP/DOWN or the lift is moved
LEFT/RIGHT.
Ensure that there are end stops installed at each end of the track.
B) Monthly - To be completed by User
Should any of the these items fail the inspection do not use the lift.
Contact your local authorized dealer for service.
Complete the visual inspection as noted in the “Each Use” section above.
With no one in the sling nor attached to the lift check the following:
The lift moves freely along the entire length of the track.
C) Semi-Annual or Yearly - To be completed by a lift technician
Consult your local authorized dealer for advice on whether this section should be completed every 6
months or on a yearly basis. Generally, in frequent use, or in situations where heavier than normal
clients are lifted, or in multi-user environments such as in institutions the lift should be checked every 6
months
Complete the visual inspection as noted in the “Monthly” section above.
Complete the preventative maintenance procedure as outlined in technical manuals for the
C1000 system.
C1000 lift checked and passed. Any required repairs completed.
This section to be only completed by a qualified service technician as authorized by Prism
Medical