Attaching the airline tube to the lift
Should the gray rubber airline that connects the lift to the hand
control become disengaged from the underside of the lift it must
be re-connected in order for the lift to work properly.
The rubber airline may become disconnected for the following
reasons:
1) The lift is pulled along the track by the airline.
2) The tubing accidentally gets wrapped around an object while
a lift or transfer is being performed.
3) It is accidentally pulled out by the caregiver or the individual
being lifted.
The airline is connected to a gray rubber grommet located on the
underside of the lift. Refer to figure 3A.
Small metal ribbed pins located at the end of the airline hold the
airline to this rubber grommet in a specific manner. Therefore it is
important to make sure that the airline is connected properly.
Both the grey airline and the rubber grommet have a grey rib on
one of their sides. Line up the grey ribs together. Refer to figure
3B. When this is done then the metal ribbed pins attached to the
end of the airline can be re-inserted into the corresponding holes
in the rubber grommet on the underside of the lift. Be sure to in-
sert the pins into the grey rubber grommet sufficiently so that it is
secure. Refer to figure 3C.
Perform a brief test to ensure proper connectivity. Turn the lift
ON and OFF. Raise and lower the carry bar. For motorized trav-
erse lifts move the lift left and then right. If these functions work
correctly then the airline is properly connected.
If the lift does not work properly, check to ensure that the grey
ribs on the grey rubber grommet on the underside of the lift and
the airline tubing are lined up properly. If they are not lined up
properly, then remove the airline, line up the grey lines and then
re-insert it into the rubber grommet. Perform the test as noted in
the preceding paragraph. If there are still problems with the lift
then contact your local authorized dealer for service.
Caution: A sturdy ladder may be required in order to access the underside of the lift to re-attach
the rubber airline of the lift. Caution should be used when this is required.
Should you have any concerns or questions contact your local authorized Prism Medical dealer.
Figure 3A - Gray rubber grommet located on
underside of lift. Rubber airline is not connected.
Note grey rib on grommet.
Figure 3B - Gray rubber airline being inserted
into rubber grommet of lift. The black ribs on
both pieces are lined up. The metal ribbed pins
are on the airline.
Figure 3C - Gray rubber airline being inserted
into rubber grommet of lift. The black ribs on
both pieces are lined up. The metal ribbed pins
are on the airline.
Metal pins that get inserted into the holes
of the grommet of the lift.
Pieces connected together