Instruction Manual Automatic Wrist Blood Pressure Monitor Model No.
Table of Contents Medical Disclaimer..........................................................................2 Intended Use.................................................................................2 About Blood Pressure ......................................................................3 Measurement Method .....................................................................5 Accuracy .......................................................................................6 Precautions ............
Medical Disclaimer This manual and product are not meant as a substitute for advice provided by your doctor. You are not to use the information contained herein, or this product for diagnosing or treating a health problem or prescribing any medication. If you have or suspect that you have a medical problem, promptly consult your healthcare provider. Intended Use This device automatically measures human’s Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during deflation.
About Blood Pressure 1. What is Blood Pressure? Blood pressure is the measurement of the force of blood pushing against the walls of the arteries. Arterial blood pressure is constantly fluctuating during the course of the cardiac cycle. The highest pressure in the cycle is called the systolic blood pressure, and represents the pressure in the artery when the heart is beating. The lowest pressure is the diastolic blood pressure, and represents the pressure in the artery when the heart is at rest.
About Blood Pressure A. AHA Blood Pressure Classifications: Standards for assessment of high or low blood pressure without regard to age, have been established by the American Heart Association (AHA 2017), as shown in the chart.
Measurement Method HL158B1-BD Automatic Wrist Blood Pressure Monitor measures blood pressure and heart rate by oscillometric method, meaning the fluctuations in pressure are measured. Once the cuff is wrapped around your wrist, just turn on the monitor and inflation automatically starts. The inflation of the cuff creates pressure around the arteries inside the wrist. Within the cuff is a gauge which senses the fluctuations (oscillations) in pressure.
Accuracy HL158B1-BD Automatic Wrist Blood Pressure Monitor has been clinically tested against a scientific device called a mercury sphygmomanometer, considered the gold standard in blood pressure measurement. All HL158B1-BD Automatic Wrist Blood Pressure Monitors have performed equivalent to measurements taken with this scientific device and are within the accuracy limits prescribed by the American National Standard for Electronic or Automated Sphygmomanometers.
Precautions * Read the Instruction Manual thoroughly before measuring and keep it at hand for your reference at any time. * The device is designed for home use and not suitable for clinical use. * The patient is an intended operator, who can operate the device by himself or herself, not necessarily by a physician or operator. * This monitor is not intended for use in the MR environment. The device should not be used to either self-diagnose Hypertension or exclude the diagnosis of Hypertension.
Precautions This product is not suitable for: Pregnant women People with arrhythmias Undergoing intravenous injection on any limb Currently in a dialysis treatment In pre-eclampsia condition For those who have had a mastectomy or lymph node clearance, it is recommended to take a measurement on the unaffected side. When used among medical electronic equipment on the same limb, pressurization of the cuff may cause temporarily malfunction to other devices.
Device Overview Product Components LCD Display Risk Category Indicator Mode Button Set/User Button Wrist Cuff Memory/ Bluetooth Button Start/Stop Button AAA “LR03” (1.
Device Overview Unit Display Hour:Minute (Also Represents Year in Setting Mode) Month/Date Systolic Pressure User Number Bluetooth Symbol Low Battery Symbol Average of Last 3 Measurements 1. Excessive Body Motion Detector AM/PM Last 3 Averaging 2.
Symbol Definitions SYMBOLS Definitions This symbol appears when the battery power is excessively low or the polarity reverses. Low Battery Symbol We suggest you replace all batteries with new ones, and make sure the +/- polarities are properly positioned. Once pulse is detected, the symbol flashes with each pulse beat. Pulse Symbol Our suggestion: Please do not talk or move during measurements. This symbol appears when an irregular heart beat was detected.
Features Risk Category Indicator (AHA 2017) This device is equipped with Risk Category Indicator which classifies your blood pressure measurements into five stages (Normal, Elevated, Hypertension stage 1, Hypertension stage 2 and Hypertensive crisis) based on the blood pressure standards established by the American Heart Association (AHA). Besides, for yours and your loved ones’ health, we further classify the five stages into numeral ranges, which sorts out hypertension symptoms more clearly.
Features Irregular Heartbeat Detector The symbol will appear on screen indicating a certain heartbeat irregularity was detected during measurement. The heartbeat rhythm that is more than or less than 25% from the average rhythm is usually defined as an irregular heartbeat rhythm. Talking, moving, shaking or an irregular pulse during the measurement can result in the appearance of this symbol.
Features HL158B1-BD has a built-in Wrist Position Guide function used as an aid to help user determining if the device is at appropriate height. Turning Wrist Position Guide ON/OFF This function default setting is on. User can switch the function ON and OFF. Under standby mode, users can press button, the device will enter to set year, date, month, hour, minute and Bluetooth function first.
Bluetooth Transmission HL158B1-BD features a Bluetooth transmission function, which enables the device automatically transmit measured results to paired Bluetooth-enabled device after measurement. When connection established, the device would transmit, systolic pressure, diastolic pressure, and pulse with time to the Bluetooth enabled device.
Bluetooth Transmission 2. If HL158B1-BD is connected successfully to your smart device, Bluetooth symbol will appear on the screen. 3. If the monitor cannot be connected to paired Bluetooth-enabled device over 45 seconds, LCD will display Error message “E4” and Bluetooth will be turned off. App for Bluetooth Download and install “DailyChek®” app on your smart device from Google Play or App store. System requirement of the Bluetooth-enabled device □ Bluetooth 4.2 for Android 6.0 or above □ Bluetooth 4.
Installing Batteries When LOW BATTERY SYMBOL appears on the display, or no reaction toward operation, please change batteries. Replace all worn-out batteries with new ones and do not mix new and used batteries. Do not mix alkaline, standard (carbon-zinc) or rechargeable (cadmium) batteries either. Such action may shorten the battery life or cause the device to malfunction. All batteries used must be the same type. Slide the battery cover, prepare 2 AAA (1.5V, LR03) alkaline batteries.
Applying the Cuff Do not place the pressure cuff over a jacket or sweater sleeve. Wrap the pressure cuff around the bare wrist with the monitor facing you. Wrap the cuff snugly. Do not make it too tight. Fold the remaining part of the cuff back out of the way. Leave approximately 0.4 inch (10 mm) between the cuff and the bottom of your hand palm. 0.4 inch (10 mm) 5.3 ~ 7.7 inch (135 ~ 195 mm) *Note! Do not use this device if your wrist has any wound or injury.
Positioning Guide It is extremely important that the cuff be at the same height as the heart. Having the cuff higher or lower may cause inaccurate results. 1. Sit on a chair comfortably, put your feet flat on the floor and lay your forearm on the table, make sure your back and arm supported, legs uncrossed. 2. Position the blood pressure monitor on your wrist. 3. Place your elbow on the table and rest the back of your hand on the device storage case or other object. 4.
Measurement Procedure Switch on the Monitor A) Put in 2 AAA 1.5V (LR03) alkaline batteries. B) All segments appear on the screen for 3 seconds. C) The monitor will automatically turn to sleeping mode (all LCD segment cleared). Setting Year, Date and Time Year, date and time can be set by two methods, either set manually using the and buttons or sync automatically using your Bluetooth smart device. 1. Set Manually A) Under standby mode, press button to enter setting mode, then YEAR digit flashes.
Measurement Procedure Taking a Measurement A. Press button to select User 1 ( ) or User 2 ( ). B. Start a Measurement: B-1. Measurement with Wrist Position Guide Off With the cuff wrapped around your wrist, press START/STOP button to start measurement. All display units appear on the screen for 3 seconds. B-2. Measurement with Wrist Position Guide On B-2-1. With the cuff wrapped around your wrist, press START/STOP button to start measurement. All display units appear on the screen for 3 seconds.
Measurement Procedure B-2-3. Even if the monitor is not in the ideal position after 10 seconds, the measurement will still begin. If this occurs, the wrist symbol will appear and be stored in the memory with the measurement. C. As the cuff inflates, the monitor automatically determines your ideal level. This monitor detects your blood pressure and pulse rate during inflation. The Pulse ( ) Symbol flashes at every pulse beat. Remain still and do not move until the entire measurement process is completed.
Memory Function Storing Data After each measurement, your systolic, diastolic, heart rate and corresponding Risk Category Indicator, Irregular Heartbeat Detector, Excessive body motion detector and wrist position guide (if any) with the time and date will be automatically stored. The monitor features 2 user memory capabilities. Each user holds the last 120 measurements, and automatically replacing the oldest data with new one.
Memory Function Recalling Data A. Press button to select User 1 or User 2. B. Press button to enter Memory Mode. If there is no data stored before, nothing (except month, date, and time) will appear on the display. If yes, the first reading will be the average of last 3 measurements. C. Press button again, LCD displays average of latest 3 AM measurements, and then is average of latest 3 PM measurements. If fewer than three measurements will not be displayed. D.
Storage and Maintenance General Use Do not twist the cuff in any way. Do not press START/STOP button if the cuff is not wrapped around your wrist. Do not drop the product and avoid any strong impacts. Maintenance Do not attempt to disassemble or change any parts of the monitor, Only trained technicians are allowed to repair and disassemble the device, including the cuff and patches, because a substitution of a component different from supplied might result in measurement error.
Storage and Maintenance According to the use environment of the sphygmomanometer, the recommended disinfection method and frequency are as follows: Only use it yourself (home use), it can be cleaned at ordinary times, and wipe it once a month with a commercially available 75% alcohol cotton sheet (for the cuff) for more than 30 seconds each time. If it is used for more than one person (home use), it can be cleaned at ordinary times.
Troubleshooting SYMBOLS/SYMPTOMS CONDITIONS/CAUSES Unit does not turn on when Worn-out batteries. START/STOP button is pushed. Battery polarities have been positioned incorrectly. INDICATION/CORRECTION Replace them with 2 new AAA (LR03) alkaline batteries. Re-insert the batteries in the correct positions. Measuring Error Symbol Cuff has been placed appears when blood incorrectly. pressure value displayed is excessively low or high. Air circuit abnormality.
Troubleshooting SYMBOLS/SYMPTOMS Cuff Fitness Detection Symbol CONDITIONS/CAUSES The cuff was wrapped incorrectly (for example too loosely or too tightly). INDICATION/CORRECTION Please reference “applying the Cuff” section to wrap the cuff correctly. Please re-pairing the BPM and Bluetooth -enabled device with each other. See the “the Bluetooth Bluetooth function is not turn on. Transmission” section to turn on Bluetooth function.
Warranty & Recalibration Warranty for Two Years from the Manufacturing Date Please note that this warranty does not cover damage caused by misuse or abuse; accident; the attachment of any unauthorized accessory; alteration to the product; improper installation; unauthorized repairs or modifications; improper use of electrical/power supply; loss of power; dropped product; malfunction or damage of an operating part from failure to provide manufacturer’s recommended maintenance; transportation damage; thef
Specifications Model Number HL158B1-BD Measurement Method Oscillometric Rated Range of Cuff Pressure 0~300 mmHg Rated Range of Determination 40~280 mmHg Measurement Range of Heart Rate Pulse:40 ~ 199 Beats / Minute Accuracy Pressure: ± 3 mmHg Pulse: ± 5 % Max. Inflation Automatic Inflation (Air Pump) Deflation Automatic (Passive Exhaust Valve) Display Liquid Crystal Display Memory 240 Memory Total for 2 Users Unit Weight 73.79mmx 60.15mmx 22.94mm (L x W x H) 2.91 x 2.37 x 0.
Note Explanation of Symbols: Symbol Explanation Health & Life Information Refer to instruction manual/booklet - TYPE BF Applied Part - To avoid inaccurate results caused by electromagnetic interference Waste of electrical and electronic equipment (WEEE) Serial number Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30cm (12 inches) to any part of the device, Otherwise, degradation of the performance of
Note *Note! This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
Appendix Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments: Emissions test RF emissions CISPR 11 Compliance Group 1 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3 Class B Not Applicable Not Applicable Electromagnetic environment – guidance RF energy is used only to maintain device’s operation.
Appendix Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments: IEC 60601 test level Immunity Compliance level Electromagnetic environment guidance – test Conducted RF IEC 61000-4-6 3V rms At 0.
Appendix Test specifications for enclosure port immunity to RF wireless communications equipment.
Blood Pressure Diary Date: □Before □After Time: Systolic / Diastolic: Date: Pulse: □Before □After Time: Systolic / Diastolic: Date: □Before □After Time: □Before □After Time: □Before □After Time: □Before □After Time: □Before □After Time: □Before □After Time: Meal Pulse: □Before □After Time: Systolic / Diastolic: Date: Meal Pulse: Systolic / Diastolic: Date: Meal Pulse: Systolic / Diastolic: Date: Meal Pulse: Systolic / Diastolic: Date: Meal Pulse: Systolic / Diastolic: Dat
P/N:323103672 VER: A001 20220325