User's Manual

General Use
Do not in any way twist the cuff.
Do not press button if the cuff is not wrapped around
your upper arm.
Do not drop the product and avoid any strong impacts.
Maintenance
Use a piece of cloth with water or mild cleansing agent to wipe the
device and dry it immediately with a dry cloth.
Do not use detergent or any strong chemicals to clean the device.
Use only a dry cloth to wipe the cuff.
Do not attempt to disassemble or change any parts of the monitor,
including arm cuff, due to substitution of a component different
from that supplied might result in measurement error.
If any suggestion or service is requested, please consult your
service station.
Storage
If the device is not to be used for a long time, please remove the
batteries from the device (leaking of battery acid can cause the
device to malfunction).
Always store the unit in the storage case after use.
Do not place the device directly under sunlight, in high
temperature, or in humid or dusty places.
Do not store the device in extremely low (less than 13 /–25 )
and high (more than 158 /70 ) temperature, nor in a place its
humidity exceeds 93% R.H.
Troubleshooting
Storage and Maintenance
SYMBOLS/SYMPTOMS
CONDITIONS/CAUSES
INDICATION/
CORRECTION
Unit does not turn on when
button is pushed.
Worn-out batteries.
Replace them with 4
new AAA (LR03)
alkaline batteries
Battery polarities have been
positioned incorrectly.
Re-insert the
batteries in the
correct positions.
Measuring Error Symbol
appears when blood
pressure value displayed is
excessively low or high.
Cuff has been placed incorrectly.
Wrap the cuff
properly so that it is
positioned correctly.
Did you talk or move during
measurement?
Measure again. Keep
arm steady during
measurement.
Shaking of the arm with the cuff
on.
Measuring Error Symbol
Air circuit abnormality. Cuff tube
may not be plugged into monitor
correctly.
Check cuff
connection. Measure
again.
Measuring Error Symbol
Inflation pressure exceeding 300
mmHg.
Switch the unit off,
then measure
again.
Measuring Error Symbol
Error determining measurement
data.
Measure again.
BPM cannot communicate
with Bluetooth device
Paring has not been completed.
Please re-pairing the
BPM and Bluetooth
device with each other.
Bluetooth function is not turn on.
Please refer to
Page 17
“Measurement
Procedure” and Page
20 “Bluetooth
Communication” to
turn on the Bluetooth
function.
The distance between BPM and
Bluetooth device is out of
transmitting range.
Please make sure the
acceptable distance
(10 meters) with
each other.
Use an incompatible Bluetooth
device.
Please refer to
Page 13 “Bluetooth
compatibility” &
Page 26
"RF Specification”
Use non-Bluetooth device.
Unexpected loss of
electrical/mechanical integrity.
Re-insert the batteries
and try again.
Return the device to
your local distributor
or importer.
Note: If "EP" appears on the display, just return the device to your local distributor or
importer.