User's Manual

BP Category Indicator
This device is equipped with BP Category Indicator which classified
the blood pressure results with WHO (World Health Organization) BP
Classifications, which are Sever Hypertension, Moderate
Hypertension, Mild Hypertension, High Normal, Normal, and Optimal.
The Corresponding LCD segment will be turned on along with the
systolic, diastolic, and pulse rate information.
Each six segments of the bar indicator corresponds to the WHO blood
pressure classification described on the below:
WHO Blood Pressure Classification
* Reference Material: Jourmal of Hypertension 1999, Vol 17 No. 2.
*Note!
When a person’s systolic and diastolic pressures fall into different categories, the higher
category should apply. Examples listed as below:
Moderate Hypertension
High Normal
Normal
Optimal
Stages of Blood Pressure Levels
Systolic
(mmHg)
Diastolic
(mmHg)
Grade 3
Severe Hypertension
180
110
Grade 2
Moderate Hypertension
160 ~ 179
100 ~ 109
Grade 1
Mild Hypertension
140 ~ 159
90 ~ 99
High-Normal
130 ~ 139
85 ~ 89
Normal
120 ~ 129
80 ~ 84
Optimal
< 120
< 80
*Note!
Please note that other risk factors (e.g. diabetes, obesity, smoking, etc.) need to be
taken into consideration and may affect these figures. Consult with your physician for
an accurate assessment.
Features
Symbol Definitions
SYMBOLS Definitions
Low Battery Symbol
This symbol appears when the battery power is excessively
low or the polarity reverses.
We suggest you replace all batteries with new ones, and
make sure the +/- polarities are properly positioned.
Pulse Symbol
Once pulse is detected, the symbol flashes with each pulse
beat.
Our suggestion:
Please do not talk or move during measurements.
Irregular Heartbeat
Detector
This symbol appears for 1 minute when the user was talking,
moving, shaking, or an irregular heart beat was detected
during measurements.
Our suggestion:
Please do not talk or move during measurements.
Repeat the measurement after resting for at least 5 minutes,
and restart your measurement while sitting down comfortably
and quietly.
Bluetooth Symbol
LCD displays this symbol when Bluetooth Function turns ON.
Risk Category Indicator
Bar
The arrowhead points out the specific Risk Category that your
measurement reading fits in.
11 12