User guide

Carefully remove the scale platform and the platter from the shipping carton. Be sure
to retain the shipping carton and all packing material in case reshipment is required.
When choosing a location for the scale platform, keep in mind that the spot should be
on a sturdy horizontal surface that is free of vibration and air currents. Vibration from
heavy equipment or appliances and air currents from fans, heating ducts, etc. . can
cause bounce on the scale display.
Place the scale platform in its desired location. Level the platform by adjusting the
four corner feet until the bubble in the bubble level, located beneath the top platter of
the platform, is within the center circle. After a level condition is achieved, test for a
stable condition by trying to rock the platform backward and forward and side to side.
Adjust the feet for final levelness and stability.
3. ELECTRICAL SPECIFICATIONS
The following chart illustrates the load cell output based on the capacity of the scale
platform (in lb) and load cell capacity (in kg). The load cell output is based on a
nominal output of 2mV/V at rated load cell capacity (i.e. if 30kg is placed on a 30kg
load cell it will have an output of approximately 2mV/V).
Model Size Capacity Load Cell Signal output at cap.
(in) (lb) (kg) (mV/V)
DXL7002 10x10 2 3 0.6
DXL7005 10x10 5 5 0.9
DXL7010 10x10 10 10 0.9
DXL7025 10x10 25 15 1.5
DXL7050 10x10 50 30 1.5
DXL8050 12x12 50 30 1.5
DXL8100 12x12 100 75 1.2
DXL9050 15x15 50 30 1.5
DXL9100 15x15 100 75 1.2
DXL9200 15x15 200 100 1.8