User Manual

C HAPTER 3:CALIBRATION
29
9. Key in the full scale capacity for the load cell(s) and press [ENTER].
For the Model 460, use the [PRINT] and [UNITS] keys to enter
numeric values. Refer to Table 2: 460 Key Functions on page 8.
- or -
Press [ENTER] to accept the displayed value.
10. The display briefly shows Updtg Gains as it updates the gain values,
then prompts CurWt Zero?.
11. Press [ENTER] to establish the current input signal as the zero
reference.
- or -
Press [SELECT] to display Zero=0mVv?. Press [ENTER] to use a
0mV/V output as the zero reference.
- or -
Press [SELECT] to display Keyin CurWt. Key in the known gross
weight already applied to the scale and press [ENTER].
- or -
Press [CLR] to bypass the zeroing option.
12. The display shows CAL OK? suggesting that the calibration is
acceptable.
Accept the calibration by pressing [ENTER] at the CAL OK? prompt.
- or -
Repeat the calibration by pressing [CLR] at the CAL OK? prompt.
13. Once the calibration is accepted, press [ENTER] at the ENTER=SAVE
prompt and again at the ENTER=EXIT prompt to save the new
calibration and exit the calibration mode.
- or -
To exit the calibration mode without
saving the new calibration, press
[CLR] at the ENTER=SAVE prompt. Then press [ENTER] at the
ENTER=UNDO
prompt and again at the ENTER=EXIT prompt to
exit the calibration mode.
SAVE CALIBRATION
The indicator will perform the calibration, display the value of the
calibration weight, and display the prompt, CAL OK?. At this point, you can