MISCELLANEOUS MESSAGES
EntryError
This error message is the most commonly used.
The primary causes are entering a value
preceding a key (such as [ZERO]) which is not
allowed, entering alpha data for a numeric
selection, or entering a fractional value for an
entry which only accepts whole numbers.
Occurs in Setup Mode or an operational mode.
WhichTx#?
The instrument is setup with more than one
custom transmit with parameter P991 set for
Prmpt (Prompt) and the [PRINT] key is
pressed. The WhichTx#? message is asking for
a custom transmit number to be entered. Key in
the custom transmit, then press [ENTER]. I.E.
[2] [ENTER], for custom transmit number 2.
COMMUNICATIONS ERROR MESSAGES
prtyXerror
This indicates that the parity of a received
character did not match the parity specified in
the Setup Mode, parameter P202. This could
also result if the baud rate (P200) or the number
of data bits (P201) is incorrect.
ovrnXerror
This indicates an over-run error where an
additional character was received while the
receive buffer of the 660 was full, and thus the
extra received character will be lost.
frmgXerror
This indicates that the stop bit of a received
character did not occur when it was expected.
This could be the result of an incorrect baud rate
(P200), incorrect number of data bits (P201), or
incorrect parity setting (P202).
portX error
The 660 did not check its receive data register in
time, thus missing a character. If this error
should occur, please notify your GSE distributor