Medical Archive Solutions User Guide

85
HP Medical Archive
Services and Components
The grid can present multiple AE titles, so associations must also
match the caller to an accepted grid AE. This is used to assign differing
profiles to single remote AEs.
DICOM Advanced
Each remote entity that supports DICOM can be allocated a profile
defining the capabilities the grid offers to that entity. This component
defines the profiles that can be assigned to AEs in the DICOM compo-
nent (above).
The Advanced Config Profile is used during the association handshake
to control the types of actions permitted and the format of files in the
Table 24: CMN
X DICOM Component Attributes
Attribute Type / Units Description
C T Alarm
DICOM AEs
Description Text User-defined description of the
entity or group of entities.
N/A
AE Title Text The AE Title of the remote entity. N/A
IP Range Dotted decimal
or CIDR
A remote device IP address (or
inclusive range) to which the grid
assigns the profile.
N/A
Port Number The port used for DICOM by the
remote device.
N/A
Via LDR Check box Indicates whether the LDR can
make direct connections with
remote entities. If not selected,
connections are routed through a
CLB service. The HP Medical
Archive must use the CLB (check
box is deselected).
N/A
GRID AE Text AE Title of the grid to which the
entity can associate.
N/A
Profile Name Text Case sensitive reference to an
entry defined in the DICOM
Advanced component. The profile
governs the permitted activities
for entities in this IP range.
N/A