Medical Archive Solutions User Guide

177
HP Medical Archive
Troubleshooting Alarms
DEAE Count of inbound DICOM C-Echo
failures.
50 is considered a minor issue.
These alarms indicate a problem
with executing specific activities
in the DICOM protocol.
Check the log at the remote entity
to determine if the errors are origi-
nating with the modality or the
grid.
If the modality shows the errors,
then adjust the configuration of
the entity.
If the errors are on the grid side of
the connection, contact HP techni-
cal support.
To reset the counters, use the
Configuration
X Main page; see
“DICOM” on page 166.
DEAF Count of inbound DICOM C-Find
failures.
50 is considered a minor issue.
DEAI Count of inbound DICOM C-Store
failures.
50 is considered a minor issue.
DEAM Count of inbound DICOM C-Move
failures.
50 is considered a minor issue.
DEAO Count of outbound DICOM C-Store
failures.
50 is considered a minor issue.
DEEO Count of outbound DICOM C-Echo
failures.
50 is considered a minor issue.
DEIA Count of inbound DICOM association
failures from external sources (after the
TCP/IP layer).
50 is considered a minor issue.
DEOA Count of outbound DICOM association
failures to external destinations (after the
TCP/IP layer).
50 is considered a minor issue.
DESC Count of inbound Storage Commitment
failures.
50 is considered a minor issue.
Attr.
Code